0.4 Micron Filter MF01-0071

adv03_advantage_filter

Product Code: ADV03

MFG Product ID : MF01-0071

( Each )



Specifications

Advantage Plus

0.4 Micron Water Pre-Filter: Compatible with Medivators Advantage Plus

 

ADV03 is a compatible filter for OEM Part MF01-0071.

Water filters should be replaced at a minimum every three months for the 1.0 micron and 0.4 micron filters and every six months for the 0.1 micron filter. If the water pressure drops below 40 PSI or the difference between the gauges is greater than 5 PSI the corresponding filter needs to be replaced. The following conditions must be met before changing a filter: 1. No endoscope reprocessing cycles are running; 2. No water line disinfection procedure is running.

 

Instructions for installing Medivators Advantage Plus water filters:

1. Relieve the pressure in the filtration system by opening the bleeder valve on the filter system (rotate down).

2. Check the inlet pressure gauge to ensure that the pressure is at zero.

3. Place an empty container under the 1 micron filter housing and disconnect the Quick Connect fitting at the bottom of the housing. Allowing all of the water to empty out.

4. Using a filter bowl wrench, remove the 1 micron filter housing and discard the used filter cartridge.

5. Place a new filter in the housing, ensuring that the O-ring is in place.

6. Thread the housing onto the header and tighten by hand only.

7. Repeat the procedure to replace the 0.4 micron and 0.1 micron filters.

8. To remove the 0.1 micron filter rotate the used filter cartridge counterclockwise, pull downward, and discard.

9. To install a new 0.1 micron filter insert a new filter cartridge into the housing header and rotate clockwise until the tabs lock into place.

10. Return the bleed valve to the closed position (rotate up).

11. Slowly open the main water supply valve and check for any leaks

*The above is adapted from Medivators guides.

 

Medivators® and Advantage Plus® are registered trademarks of Medivators Inc. MedService Repair is not affiliated with Medivators and any and all references to Medivators® on this website are intended to be for informational purposes only. The products on this site have not been manufactured or approved by Medivators®.

MedService Repair is your #1 source for OEM compatible AER filters at the lowest possible prices. We have years of experience with AER filters! We’ve been servicing and repairing AERs for decades which gives us firsthand knowledge of what works and what doesn’t. That knowledge becomes your advantage as we’ve spent a great deal of time considering the water filters and maintenance products we offer our clients and use ourselves repairing and maintaining AERs for medical offices throughout the United States. By creating a free account today you’ll have access to our member only special pricing on AER filters and many other parts and service items to keep your endoscope reprocessor in excellent working condition.

For years, buying the manufacturer’s OEM AER filters was the only logical solution as they were the highest quality parts. This was the was case for all the most popular AER brands; Medivators AER Filters, Cantel AER Filters, ASP AER Filters, Johnson & Johnson AER Filters, Olympus AER Filters, Custom Ultrasonics AER Filters, Steris AER Filters.

And while OEM quality has not suffered, today other options exist. Manufacturer compatible AER filters are available from reputable sources you can trust at highly discounted prices. The savings available is instrumental to running medical practices efficiently especially when the quality of the AER replacement and maintenance parts themselves meets or exceeds the manufacturer’s own high quality water filters.

The manufacturing processes of AER filters have evolved considerably over the years. Automation and technology advances in the manufacturing processes play a large role in this. For instance, the quality of the side seam welding has improved dramatically with new equipment that has helped these competing AER filter manufacturers meet or exceed the OEM manufacturer’s own strict guidelines. The glues and additives that used to be used and were a source of common issues have been replaced with ultrasonic thermal welding which completely ensured the integrity of the seams.

All this to say that if you haven’t considered aftermarket OEM compatible AER Filters as the solution for your medical practice’s consumable maintenance items you may want to consider it. If you have any questions at all about AER Filters, please call us today at the number below. We’ll be happy to answer any questions and make recommendations.

 

Wassenburg Medical Acquires Custom Ultrasonics

On December 31st, 2021 Wassenburg Medical purchased the assets of Custom Ultrasonics and will assume the activities of Custom Ultrasonics.

Custom Ultrasonics was established in 1979 by Frank J. Weber focusing on the healthcare industry with regard to infection control.  In 1981 they introduced the System 83. The System 83 is an automated system to clean and disinfect multichannel flexible endoscopes.

Wassenburg Medical serves all healthcare entities including Hospitals, Medical Centers, Outpatient Clinics, Ambulatory Surgery Centers, physician offices and any healthcare facility where endoscopy procedures are performed. Wassenburg Medical specializes in effective cleaning and high-level disinfection.